PublicCVE

CVE-2024-48970

CRITICAL9.3

Description

The ventilator's microcontroller lacks memory protection. An attacker could connect to the internal JTAG interface and read or write to flash memory using an off-the-shelf debugging tool, which could disrupt the function of the device and/or cause unauthorized information disclosure.

CVSS breakdown

CVSS 3.1
Attack Vector
Local
Attack Complexity
Low
Privileges Required
None
User Interaction
None
Scope
Changed
Confidentiality
High
Integrity
High
Availability
High

Affected products