Description
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.
CVSS breakdown
CVSS 3.1
Attack Vector
Local
Attack Complexity
Low
Privileges Required
None
User Interaction
None
Scope
Changed
Confidentiality
High
Integrity
High
Availability
High
Affected products
- Baxter / Life2000 Ventilation System06.08.00.00 and prior – 06.08.00.00 and prior